Pooled sample testing for COVID-19 diagnosis: Evaluation of bi-directional matrix pooling strategies
扫码查看
点击上方二维码区域,可以放大扫码查看
原文链接
NSTL
Elsevier
In the on-going COVID-19 pandemic, pooled testing of samples by RT-PCR has been recommended at certain scenarios to increase labs' testing capacity and reduce cost of testing. This paper describes the evaluation of bidirectional matrix pooling strategies with clinical samples in a 5 x 5 and 10 x 10 matrix. Nasopharyngeal swab samples in viral transport medium (VTM) previously tested (positive or negative) by real time RT-PCR for SARSCoV-2 were used for these experiments. Ten sets of 5 x 5 (250 samples) and ten sets of 10 x 10 (1000 samples) pooling of samples in both directions was done with known positive samples introduced at random positions. Extracted nucleic acid was tested for SARS-CoV-2 E-gene by RT-PCR. Sensitivity or concordance and feasibility of matrix pooling were assessed in comparison to direct RT-PCR testing. In comparison to direct testing, the overall concordance was 86.6% for 5 x 5 pooling, 73.3% for 10 x 10 with 200 mu L extraction volume and 86.6% for 10 x 10 with 400 mu L extraction volume. Bi-directional matrix pooling can be adopted with advantage over conventional direct or pool testing for COVID-19 by RT-PCR under the following conditions: i) sample positivity rate of < 5%, ii) matrix pool size of 8-10 samples, iii) use of min. 40 mu L VTM from each sample and iv) utilization of automated liquid handling equipment, if available, for sample addition to avoid human errors.