Substantial peak size effect on compound-specific delta D values analyzed on isotope ratio mass spectrometry
扫码查看
点击上方二维码区域,可以放大扫码查看
原文链接
NSTL
Elsevier
Lipid compound-specific hydrogen isotopes (delta D) have been widely applied in studies of biogeochemistry and paleoclimatology. delta D values of lipid homologues with different chain lengths from a single analysis are often utilized together in order to make full use of the delta D information. However, this approach could be potentially impacted by the large analytical error due to the inappropriate amount injected for some homologues. Here, we systematically investigate compound-specific delta D deviations with varying contents injected on the isotope ratio mass spectrometry, expressed as amplitudes, using both lab working standards and natural samples. The n-alkane delta D values of lab working standards (mixed C-21, C-25, C-27, C-29, C-31, and C-33) could vary by 40 parts per thousand to 70 parts per thousand when the amplitudes of n-alkanes change from 0.5 V to 10 V. For natural samples (27 n-alkane and 77 fatty acid samples), we have made repeated analyses of the same samples with different homologues targeted for the optimum range. The measured delta D values are higher by 20 parts per thousand to 40 parts per thousand with the amplitudes lower than the optimum range, and lower by 10 parts per thousand to 20 parts per thousand with higher amplitudes. All the results consistently show higher delta D values with decreasing amplitudes, and larger deviations occurring in low amplitude range, implying that special caution should be taken with the delta D values measured at low amplitude range. We have attempted amplitude-based correction of lipid delta D values, however, this approach should be cautious owing to their large residual errors. Hence, delta D values of different homologues in the same samples have to be measured in separate analysis if they could not fall within the optimum range at the same time.
Hydrogen isotopesN-AlkanesFatty acidsIsotope ratio mass spectrometryN-ALKANE DISTRIBUTIONSPLANT LEAF WAXESLIPID BIOMARKERSTERRESTRIAL PLANTSSUBMERGED PLANTSFATTY-ACIDSD/H RATIOSHYDROGENSEDIMENTSVARIABILITY