Cryogenics2022,Vol.1236.DOI:10.1016/j.cryogenics.2022.103451

Strain measurement by means of clip-on extensometers during discontinuous plastic flow at 4 K

Tabin, Jakub
Cryogenics2022,Vol.1236.DOI:10.1016/j.cryogenics.2022.103451

Strain measurement by means of clip-on extensometers during discontinuous plastic flow at 4 K

Tabin, Jakub1
扫码查看

作者信息

  • 1. Cracow Univ Technol
  • 折叠

Abstract

The clip-on extensometers are recommended by international standards (ASTM E 1450-16 or ISO 6892-4:2015) for tension testing of structural alloys at 4 K. They are also frequently used to identify mechanical and electromechanical properties of superconducting wires or cables at cryogenic temperatures. In this study, the time responses of clip-on extensometers during uniaxial tensile tests of the engineering materials at liquid helium temperature are analysed. It turns out, that strong strain localization and self-excited vibrations of extensometers related to the plastic flow instability disturb the strain measurement. These measurement artefacts may affect the identification of the mechanical parameters, such as the 0.2% proof stress or the total elongation.

Key words

Clip-on extensometers/Discontinuous plastic flow/Strain localization/Cryogenic temperatures/International standards/STRUCTURAL-MATERIALS/ALLOYS/METALS/TRANSFORMATION/TEMPERATURES/DEFORMATION/MECHANISM/BEHAVIOR/STEEL