18F-NaF与99Tcm-MDP全身骨显像图像质量与辐射剂量率的对比研究

扫码查看
目的 对比研究18F-NaF PET全身骨显像与99Tcm-MDP SPECT全身骨显像图像质量和辐射剂量率.方法 收集2021年10月至2022年1月在桂林医学院附属核医学科行全身骨显像的患者97例,分别行99Tcm-MDP和18F-NaF显像.收集患者相关临床资料、注射后检查等待时间、检查结束后及24h的1m距离的平均辐射剂量率及诊断医师对两种显像的图像质量,并对两种显像剂患者组间差异进行比较.结果 18F-NaF全身骨显像注射后检查等待时间明显少于99Tcm-MDP全身骨显像,且差异具有统计学意义(P<0.001);18F-NaF全身骨显像图像质量高于99Tcm-MDP全身骨显像,图像质量评分(2.8±0.4 vs 2.1±0.7,t=-6.767,P<0.001),且差异有统计学意义.检查后18F-NaF全身骨显像对周围环境的辐射剂量率高于99Tcm-MDP全身骨显像,且差异具有统计学意义(17.2±6.3μSv/h vs 5.7±4.9μSv/h,t=-9.278,P<0.001).而24h后两显像方式的辐射剂量率相似,且差异不具有统计学意义(0.5±0.3μSv/h vs 0.6±0.3μSv/h,t=1.653,P=0.102).结论 18F-NaF全身骨显像较99Tcm-MDP全身骨显像缩短了检查等待时间,并具有更高的图像质量和评分一致性.另外,两种显像剂检查后24h辐射剂量率相似.
A Comparative Study of Image Quality and Residual Dose Between 18F-NaF and 99Tcm-MDP Whole Body Bone Imaging
Objective To compare the image quality and radiation dose rate of 18F-NaF PET whole body bone imaging and 99Tcm-MDP SPECT whole body bone imaging.iMethods From October,2021 to January,2022,97 patients who underwent whole-body bone imaging in the Nuclear Medicine Department of Guilin Hospital were collected for the study.99Tcm-MDP and 18F-NaF imaging were performed respectively.Clinical data of all patients,as well as post-injection examination waiting time,average radiation dose rate at 1m distance after examination and 24h,and image quality between the two imaging agents were collected,and the differences between the two imaging agents were evaluated.Results The waiting time of 18F-NaF whole-body bone imaging after injection was significantly shorter than that of 99Tcm-MDP whole-body bone imaging(P<0.001).The image quality of18F-NaF whole-body bone imaging was significantly higher than that of 99Tcm-MDP whole-body bone imaging(2.8±0.4 vs 2.1±0.7,t=-6.767,P<0.001).After examination,the radiation dose rate of 18F-NaF whole-body bone imaging was significantly higher than that of 99Tcm-MDP whole-body bone imaging(17.2±6.3μSv/h vs 5.7±4.9μSv/h,t=-9.278,P<0.001).After 24 hours,the radiation dose rates of the two imaging methods were similar,and the difference was not statistically significant(0.5±0.3μSv/h vs 0.6±0.3μSv/h,t=1.653,P=0.102).Conclusion Compared with 99Tcm-MDP whole-body bone imaging,18F-NaF whole-body bone imaging shortens the waiting time for examination,and shows a higher image quality and consistency of scoring.In addition,the radiation dose rate of 24 hours after examination of the two imaging agents is similar.

99Tcm-Methylenediphosphonate18F-FluorideWhole-body bone imagingRadiation dose rate

莫碧云、牟兴宇、黎祖国、曾钰瀧、崇维霞、付巍

展开 >

桂林医学院附属医院核医学科,广西桂林,541000

99Tcm-亚甲基二膦酸盐 氟化钠 全身骨显像 辐射剂量率

2024

标记免疫分析与临床
中国同辐股份有限公司

标记免疫分析与临床

CSTPCD
影响因子:0.978
ISSN:1006-1703
年,卷(期):2024.31(11)