Determination of calcium ion in Calcium Acetate Granules by ion chromatography-suppressed conductivity method
Objective To establish a method of ion chromatography-suppressed conductivity to determine calcium ion in Calcium Acetate Granules.Methods The assay was performed on Dionex IonPacTM CS12A column(250 mm×4 mm)with Dionex IonPacTM CG12A column(50 mm×4 mm)as guard column.Mobile phase was used as 0.02 mol/L methanesulfonic acid solution.Flow rate was 1.0 mL/min,column temperature was 30℃,and detecting pool temperature was 35℃.Conductivity detector was used with CDRS 600 4 mm as suppressor,current value was 59 mA,and injection volume was 25 μL.Results Calcium showed good linear relationships of 0.002—6.000 μg/mL(r=0.999 7),whose average recovery was 104.6% with RSD values of 1.8%.Conclusion The method is simple,highly specific,accurate and reliable,and can effectively control calcium ion in Calcium Acetate Granules.