In the daily production and detection process of ARL4460 direct reading spectrometer,it was found that the standardization coefficient of N channel was extremely abnormal,and the test accuracy of N element was significantly reduced.After checking the possible influencing factors one by one,the outgoing light path of locking N channel was seriously offset,and finally the problem of reduced sensitivity of N element analysis was successfully solved by adjusting the light path of N channel.
Direct reading spectrometerNitrogenSensitivityOptical path