首页|无线传能尿道阀电磁安全性仿真分析

无线传能尿道阀电磁安全性仿真分析

扫码查看
无线传能尿道阀是一种模拟尿道括约肌功能而设计的辅助控尿装置,为评估它植入人体后工作的安全性,利用HFSS软件建立人体各组织层三维模型,计算无线传能线圈间距分别为 15、20 mm时、100~1 000 kHz频率内的人体各组织层SAR值与电流密度值,分析各组织层SAR值、电流密度值随无线传能线圈间距与工作频率的变化规律.仿真结果表明:各组织层中SAR值与电流密度值均随工作频率的增大而增大,皮肤层、脂肪层的SAR值与电流密度随线圈间距增大呈增大趋势,而肌肉层的SAR值与电流密度随线圈间距增大而减小;最大SAR值与电流密度值分别为 8.77×10-11 W/kg、433.5 μA/m2,均远小于安全限值,满足安全使用要求.
Electromagnetic Safety Analysis of Wireless Energy Transfer Urethral Valve
Wireless energy transfer urethral valve is a kind of auxiliary urinary control device,which is designed to simulate the function of the urethral sphincter.In order to evaluate its safety after implantation in the human body,3D models of human tissue layers were established using HFSS software.Specific absorption rate(SAR)and current density values of all tissue layers were calculated in the frequency range from 100 kHz to 1 000 kHz when the distance between wireless energy transfer coils was 15 mm and 20 mm.The changes of SAR value and current density of each tissue layer with the distance of wireless energy transmission coils and operating fre-quency were analyzed.The simulation results show that the SAR value and current density of each tissue layer increase with the increase of operating frequency,the SAR value and current density of the skin layer and fat layer increase with the increase of coil distance,while the SAR value and current density of muscle layer decrease with the increase of coil distance.The maximum SAR value and current den-sity value are 8.77×10-11 W/kg and 433.5 μA/m2,respectively,which are far less than the safety limit and meet the requirements of safe use.

urethral valvewireless energy transferelectromagnetic safetysimulation analysis

尹茂、李笑

展开 >

广东工业大学机电工程学院,广东广州 510006

尿道阀 无线传能 电磁安全性 仿真分析

国家自然科学基金面上项目广东省自然科学基金项目广州市科学研究计划

520751012017A030313265201904010184

2024

机床与液压
中国机械工程学会 广州机械科学研究院有限公司

机床与液压

CSTPCD北大核心
影响因子:0.32
ISSN:1001-3881
年,卷(期):2024.52(16)