防护热板导热仪校准热流计的研究

Research on the Calibration of Heat Flux Meters Using Guarded Hot Plate Apparatus

扫码查看
防护热板导热仪具有在计量板上产生精确可测量的单向热流的能力.为此,采用中国计量科学研究院研制的高性能单试样防护热板导热仪对2 支国外精准热流计与2 支国产热流计进行校准方法的研究,以40 mm厚岩棉作为热阻尼围护材料,设计了将热流计的测试面贴近计量板(工况1)、放置在岩棉的中间(工况2)和贴近冷板(工况3)三种工况,测量各个热流计的灵敏度系数C.工况 2 下,C的实测值与 2 支国外的热流计示值的偏差最小,相对偏差约7%;4 支热流计的实验结果一致显示,工况2 下C值最低,工况1 最高.分析认为是构成热流计的有机固体物的导热系数大于岩棉,同时接触计量板或冷板面,热流穿过热流计单元的阻力小于穿过岩棉,导致热流的分布不是单方向的,接触面上的热流分布不均匀,热流计单元吸收了更多的热流,使得C的示值偏大.
The guarded hot plate apparatus has the capability to generate precise and measurable unidirectional heat flow on the measurement plate.The calibration method of two foreign precision heat flow meters and two domestic heat flow meters was investigated using a high-performance single-sample guarded hot plate apparatus developed by National Institute of Metrology of China.A 40 mm thick rock wool was used as the thermal resistance insulation material.Three different working conditions were designed:placing the test surface of the heat flow meter close to the measuring plate(Condition 1),placing it in the middle of the rock wool(Condition 2),and attaching it to the cold plate(Condition 3).The sensitivity coefficient C of each heat flow meter was measured.Under Condition 2,the measured value of C showed the smallest deviation from the indications of the two foreign heat flow meters,with a relative deviation of about 7%.The experimental results of all four heat flow meters consistently showed that Condition 2 had the lowest C value,while Condition 1 had the highest.Analysis indicated that the thermal conductivity of the organic solid material constituting the heat flow meter was greater than that of rock wool.Additionally,when in contact with the measuring plate or the cold plate,the resistance of the heat flow passing through the heat flow meter unit was smaller than that passing through the rock wool,resulting in non-unidirectional distribution of the heat flow and uneven distribution of the heat flow on the contact surface,the heat flow meter unit absorbed more heat flow,leading to an overestimation of the C value.

thermal measurementguarded hot plate apparatusthermal resistance heat flux meterheat fluxcalibration

孙泽龙、许淑惠、张金涛

展开 >

北京建筑大学 环境与能源工程学院,北京 102616

中国计量科学研究院 热工计量科学研究所,北京 100029

热学计量 防护热板导热仪 热阻式热流计 热流 校准

2024

计量学报
中国计量测试学会

计量学报

CSTPCD北大核心
影响因子:0.303
ISSN:1000-1158
年,卷(期):2024.45(12)