科学技术与工程2024,Vol.24Issue(19) :8018-8024.DOI:10.12404/j.issn.1671-1815.2305383

菱形与非平行空心加压螺钉治疗股骨颈骨折的有限元分析

Comparison of Diamond and Nonparallel Fixation with Cannulated Screws for Femoral Neck Fracture of Pauwels:A Finite Element Analysis

李志远 南宠 田金辉 江仲超 李晓东 刘炳智 苗洁
科学技术与工程2024,Vol.24Issue(19) :8018-8024.DOI:10.12404/j.issn.1671-1815.2305383

菱形与非平行空心加压螺钉治疗股骨颈骨折的有限元分析

Comparison of Diamond and Nonparallel Fixation with Cannulated Screws for Femoral Neck Fracture of Pauwels:A Finite Element Analysis

李志远 1南宠 2田金辉 1江仲超 1李晓东 1刘炳智 1苗洁1
扫码查看

作者信息

  • 1. 邯郸市中心医院骨科,邯郸 056001
  • 2. 保定市第一中心医院骨科,保定 071000
  • 折叠

摘要

为对比分析菱形与非平行空心加压螺钉治疗3种类型(Pauwels Ⅰ、Ⅱ、Ⅲ)股骨颈骨折在模拟生物载荷下的力学差异.通过使用计算机三维处理软件(Minics、Solideworks、Geomagic)分析临床提供的CT数据进行建模,包括3种类型(PauwelsⅠ、Ⅱ、Ⅲ)及两种不同排列方式的空心加压螺钉模型;导入Ansys进行网格划分、材料赋值,模拟生理载荷,观察最大形变和内固定的等效应力和剪切应力的方法.对于菱形和非平行空心钉模型,在Pauwels Ⅰ型(20°)骨折中,最大形变0.560 2、0.5610 mm,最大等效应力23.68、24.49 MPa,最大剪切应力3.715、4.160 MPa;在Pauwels Ⅱ型(40°)骨折中,最大形变0.562 0、0.565 3 mm,最大等效应力24.64、25.32 MPa,最大剪切应力4.540、5.549 MPa;在Pauwels Ⅲ型(60°)骨折中,最大形变0.5708、0.572 3 mm,最大等效应力25.91、26.08 MPa,最大剪切应力6.668、7.089 MPa.可见在同等受力条件下,菱形空心钉模型的形变更小,内固定所承担的等效应力、剪切力更小,应力分布更为平均,更具有力学稳定性.

Abstract

In order to compare the biomechanical properties of four screws internal fixation for femoral neck fracture of Pauwels typeⅠ、Ⅱ、Ⅲ:diamond and nonparallel screws,the clinical CT images were used to establish models of three kinds of different Pauwels types and two kinds of configuration models through three-dimensional virtual software(Minics、Solideworks、Geomagic).Models were imported into software Ansys to assume material properties and load pressure.The maximum displacement,maximum stress and shear stress were observed.The biomechanical properties of diamond and nonparallel fixation with cannulated screws were differento Among Pauwels Ⅰ fracture(20°),the max displacement were 0.560 2 mm、0.561 0 mm,the max equivalent stresses were 23.68 MPa、24.49 MPa,the max shear stresses were 3.715 MPa、4.160 MPa.Among Pauwels Ⅱ fracture(40°),the max displacement were 0.562 0 mm、0.565 3 mm,the max equivalent stresses were 24.64 MPa、25.32 MPa,the max shear stresses were 4.540 MPa、5.549 MPa.Among Pauwels Ⅲ fracture(60°),the max displacement were 0.570 8 mm、0.572 3 mm,the max equivalent stresses were 25.91 MPa、26.08 MPa,the max shear stresses were 6.668 MPa、7.089 MPa.It is concluded the diamond configuration models has better stress dispersion effects,less displacement and shear stress conditions,better biomechanical stability under the same stress.

关键词

股骨颈骨折/空心钉/有限元/生物力学

Key words

femoral neck fracture/cannulated screws/finite element analysis/biomechanics