首页|多参数水质分析仪法与国标法测定水中游离余氯的比较研究

多参数水质分析仪法与国标法测定水中游离余氯的比较研究

扫码查看
目的:对多参数水质分析仪法与国标法测定水中游离余氯进行对比研究.方法:采用氯标准溶液对多参数水质分析仪的内置工作曲线进行验证,采用加标回收实验对多参数水质分析仪法进行测试.结果:多参数水质分析仪工作曲线准确可靠,回收率在80.0%~95.0%,精密度在0.38%~6.24%,符合标准要求.结论:与国标法相比,多参数水质分析仪法检测速度更快,精密度更高,适用于生活饮用水中游离余氯的快速检测.
Comparative Study on Determination of Free Residual Chlorine in Water by Multi-parameter Water Quality Analyzer Method and National Standard Method
Objective:To compare the determination of free residual chlorine in water by multi-parameter water quality analyzer method and national standard method.Method:The built-in working curve of the multi-parameter water quality analyzer was verified by chlorine standard solution,and the multi-parameter water quality analyzer method was tested by recovery test.Result:The working curve of the multi-parameter water quality analyzer was accurate and reliable,with a recovery rate of 80.0%to 95.0%and a precision of 0.38%to 6.24%,which met the standard requirements.Conclusion:Compared with the national standard method,the multi-parameter water quality analyzer method has faster detection speed and higher precision,and is suitable for the rapid detection of free residual chlorine in drinking water.

muti-parameter water quality analyzernational standard methodfree residual chlorinedrinking water

于卓然、段瑞晶、薛钦、王丽静、杜利君、薛颖、刘春光

展开 >

上海海关,上海 200000

太原海关技术中心,山西 太原 030024

山西小浪底引黄水务集团有限公司,山西 运城 044000

多参数水质分析仪 国家标准方法 游离余氯 生活饮用水

2024

现代食品
国家粮食储备局郑州科学研究设计院

现代食品

影响因子:0.169
ISSN:2096-5060
年,卷(期):2024.30(19)