多参数水质分析仪法与国标法测定水中游离余氯的比较研究
Comparative Study on Determination of Free Residual Chlorine in Water by Multi-parameter Water Quality Analyzer Method and National Standard Method
于卓然 1段瑞晶 2薛钦 3王丽静 2杜利君 2薛颖 2刘春光2
作者信息
- 1. 上海海关,上海 200000
- 2. 太原海关技术中心,山西 太原 030024
- 3. 山西小浪底引黄水务集团有限公司,山西 运城 044000
- 折叠
摘要
目的:对多参数水质分析仪法与国标法测定水中游离余氯进行对比研究.方法:采用氯标准溶液对多参数水质分析仪的内置工作曲线进行验证,采用加标回收实验对多参数水质分析仪法进行测试.结果:多参数水质分析仪工作曲线准确可靠,回收率在80.0%~95.0%,精密度在0.38%~6.24%,符合标准要求.结论:与国标法相比,多参数水质分析仪法检测速度更快,精密度更高,适用于生活饮用水中游离余氯的快速检测.
Abstract
Objective:To compare the determination of free residual chlorine in water by multi-parameter water quality analyzer method and national standard method.Method:The built-in working curve of the multi-parameter water quality analyzer was verified by chlorine standard solution,and the multi-parameter water quality analyzer method was tested by recovery test.Result:The working curve of the multi-parameter water quality analyzer was accurate and reliable,with a recovery rate of 80.0%to 95.0%and a precision of 0.38%to 6.24%,which met the standard requirements.Conclusion:Compared with the national standard method,the multi-parameter water quality analyzer method has faster detection speed and higher precision,and is suitable for the rapid detection of free residual chlorine in drinking water.
关键词
多参数水质分析仪/国家标准方法/游离余氯/生活饮用水Key words
muti-parameter water quality analyzer/national standard method/free residual chlorine/drinking water引用本文复制引用
出版年
2024