After the sample was dissolved in concentrated acid,organic matter and iodine were removed by hydrogen peroxide,and the solution was prepared at an appropriate concentration.The sample was sprayed into the air-acetylene lean flame atomization system of an atomic absorption spectrometer using a hollow cathode lamp as the light source,and the absorbance was measured at 345.3 nm wavelength.The amount of rhodium in the sample was quantified using a standard curve.After verification,the relative standard deviation of the six determinations of the same sample was less than 0.5%,the recovery rate was between 98.4%~101.8%,the detection limit was 0.027 mg/L,and the method was simple to operate and fast to determine,which could meet the requirements for the determination of rhodium content in triiodide rhodium.