IOL Master700 与Pentacam测量白内障患者角膜曲率及散光的对比研究

扫码查看
目的 探讨比较IOL Master700 和Pentacam在白内障患者术前检测角膜曲率与散光的差异性、相关性及一致性.方法 收集 2023 年 6 月至2023 年10 月白内障患者115 例203 眼.分别采用IOL Master与Pentacam测量患者的角膜前表面曲率及散光度数、后表面曲率及散光度数和全角膜曲率及散光度数,分析两种仪器的差异性、相关性和一致性.结果 IOL Master700 与Pentacam测量角膜后表面的Ks、Kf和Km的差异均具有统计学意义(P<0.05),其余结果差异均无统计学意义(P>0.05);两种仪器分别测量Ks、Kf、Km和△k的数据有较好的相关性,Ks、Kf、Km较△k相关性更高(P<0.001);Bland-Altman分析结果显示IOL Master与Pentacam测量患者的角膜前表面曲率的一致性较好,角膜后表面曲率及散光度数的一致性较差.结论 IOL Master700 与Pentacam在测量角膜前表面曲率方面差异性小,一致性好;在测量全角膜曲率方面差异性较小,一致性较好;在测量角膜后表面曲率方面存在差异,一致性较差;在测量角膜前后表面及全角膜散光度方面,差异性较小,但一致性较差.总体来说,两种仪器测量数据临床使用时不可相互替代,但可相互参考,需综合分析测量结果进行选择.
Comparative analysis of corneal keratometry and astigmatism measurements between IOL Master700 and Pentacam in cataract patients
Objective To investigate the difference and consistency of corneal keratometry and astigmatism measured by IOL Master700 and Pentacam in cataract patients.Methods A total of 115 cataract patients(203 eyes)were select-ed from June 2023 to October 2023.Corneal keratometry and astigmatism of the anterior corneal surface,posterior cor-neal surface and total cornea of cataract patients were measured by IOL Master700 and Pentacam,respectively.Differ-ences,correlations and agreement between the two instruments were analysed.Results Significant differences were found in Ks,Kf and Km of the posterior corneal surface between IOL Master700 and Pentacam(P<0.05),while no significant differences were found in other parameters.Pearson correlation analysis showed a strong correlation between the two instruments in the measurement of Ks,Kf,Km and△k.The correlations for Ks,Kf and Km were more signif-icant than those for △k(P<0.001).Bland-Altman analysis showed strong agreement between the anterior corneal sur-face Ks,Kf and Km measured by the two instruments,but poor agreement between the posterior corneal surface Ks,Kf and Km.Conclusion Firstly,the IOL Master700 and Pentacam show moderate differences and strong consistency in anterior corneal surface keratometry.Secondly,there are minimal differences and significant consistency in the measure-ment of total corneal keratometry.However,significant differences and poor consistency are observed in the posterior corneal surface keratometry.Although the differences in astigmatism measurements for the anterior corneal surface,pos-terior corneal surface and total cornea are minimal between two devices,the consistency is poor.Overall,measurement data from the IOL Master700 and Pentacam should not be used interchangeably in clinical practice,but can be cross-ref-erenced.A comprehensive analysis of the measurement results is recommended for clinical decision making.

IOL Master 700PentacamCataractKeratometryAstigmatism

杜祥阁、徐文文、刘妍、王琛、常海燕

展开 >

山东第一医科大学附属省立医院 眼科,山东 济南 250012

山东第一医科大学附属省立医院,耳鼻咽喉头颈外科,山东 济南 250012

IOL Master700 Pentacam 白内障 角膜曲率 散光

2024

山东大学学报(医学版)
山东大学

山东大学学报(医学版)

CSTPCD北大核心
影响因子:0.841
ISSN:1671-7554
年,卷(期):2024.62(12)