石化技术与应用2024,Vol.42Issue(3) :225-228.DOI:10.19909/j.cnki.ISSN1009-0045.2024.03.0225

采用差示扫描量热法测定蜡滴熔点

Determination of drop melting point of wax by differential scanning calorimetry

李淑杰 汲永钢 关旭 李奕睿
石化技术与应用2024,Vol.42Issue(3) :225-228.DOI:10.19909/j.cnki.ISSN1009-0045.2024.03.0225

采用差示扫描量热法测定蜡滴熔点

Determination of drop melting point of wax by differential scanning calorimetry

李淑杰 1汲永钢 1关旭 1李奕睿1
扫码查看

作者信息

  • 1. 中国石油石油化工研究院大庆化工研究中心,黑龙江大庆 163714
  • 折叠

摘要

采用差示扫描量热(DSC)法测定蜡滴熔点,研究了升(降)温速率、升温次数对测试结果的影响,并在最优条件下考察了 DSC法测蜡滴熔点的重复性、准确性和再现性.结果表明:确定熔融曲线终止边的拐点温度为DSC法的滴熔点;DSC法测蜡滴熔点的最优条件为:第1次升(降)温速率均为20℃/min,第2次升温速率为5℃/min;该方法的重复性在0~0.2℃,再现性在0~1.1℃,均满足GB/T 8026-2014的要求.

Abstract

The drop melting point of wax was de-termined by using differential scanning calorimetry(DSC),and the effects of heating(cooling)rate and heating times on the test results were studied.The repeatability,accuracy and reproducibility of DSC method for measuring wax drop melting point were investigated under the optimal conditions.The re-sults showed that the inflection point temperature at the end of the melting curve was determined to be the drop melting point of the DSC method.The opti-mal conditions for measuring the drop melting point of wax by DSC method were as follows:the first heat-ing(cooling)rate was 20 ℃/min,the second heating rate was 5 ℃/min.The repeatability of the method was 0-0.2 ℃,and the reproducibility was 0-1.1 ℃,which both met the requirements of GB/T 8026-2014.

关键词

差示扫描量热/滴熔点/微晶蜡/费托蜡/聚乙烯蜡/棕榈蜡

Key words

differential scanning calorimetry/drop melting point/microcrystalline wax/fischer-tropsch wax/polyethylene wax/carnauba wax

引用本文复制引用

出版年

2024
石化技术与应用
兰州石化公司石油化工研究院

石化技术与应用

CSTPCD
影响因子:0.491
ISSN:1009-0045
段落导航相关论文