医疗装备2024,Vol.37Issue(18) :51-54.DOI:10.3969/j.issn.1002-2376.2024.18.013

高温蒸汽灭菌对耳鼻喉科器械的消毒效果及循环疲劳性能的影响

The Effect of Sterilization by High-Temperature Steam on the Disinfection Efficiency and Cyclic Fatigue Performance of Otolaryngological Instruments

杨能 林芬芬 许丹慧
医疗装备2024,Vol.37Issue(18) :51-54.DOI:10.3969/j.issn.1002-2376.2024.18.013

高温蒸汽灭菌对耳鼻喉科器械的消毒效果及循环疲劳性能的影响

The Effect of Sterilization by High-Temperature Steam on the Disinfection Efficiency and Cyclic Fatigue Performance of Otolaryngological Instruments

杨能 1林芬芬 1许丹慧1
扫码查看

作者信息

  • 1. 浙江省台州医院 (浙江台州 317000)
  • 折叠

摘要

目的 探讨高温蒸汽灭菌对耳鼻喉科器械的消毒效果及循环疲劳性能的影响.方法 随机选取医院耳鼻喉科的 260 件医疗器械研究对象,按照消毒方法的不同分为常规组和高温蒸汽灭菌组,每组 130 件.常规组采用常规擦拭消毒法,高温蒸汽灭菌组使用高温蒸汽灭菌法.对比两组器械的消毒效果、循环疲劳性能、断口形貌观察、灭菌合格率.结果 高温蒸汽灭菌组器械的菌落数低于常规组,差异有统计学意义(P<0.05);高温蒸汽灭菌组器械的断裂循环次数(NCF)与器具断端的长度(FL)均高于常规组,差异有统计学意义(P<0.05);高温蒸汽灭菌组器械显微镜下断口主要表现为产生韧性开裂,伴有脆性延晶体开裂,晶界洁净且平滑,呈现片状脆性沉淀相,多面立体感强烈,棱角分明.常规组显微镜下断口表现为韧性开裂,断裂处可见明显微孔洞,断口表面覆盖了大量韧窝结构.结论 高温蒸汽灭菌的消毒效果较好,但长期使用对耳鼻喉科器械造成的损害大于常规擦拭消毒法.

Abstract

Objective To explore the effect of sterilization by high-temperature steam on the disinfection efficiency and cyclic fatigue performance of otolaryngological instruments.Methods With the random selection of 260 medical devices commonly used in the otolaryngology department of the hospital as research subjects from January 2023 to January 2024,according to different disinfection methods,the instruments were divided into the conventional group and the high-temperature steam sterilization group,with 130 pieces in each group.With the use of conventional wiping disinfection method in the conventional group and high-temperature steam sterilization method in the high-temperature steam sterilization group,the disinfection effect,cyclic fatigue performance,fracture morphology observation,and sterilization qualification rate between the two groups of instruments were compared.Results The bacterial count of the instruments in the high-temperature steam sterilization group was lower than that in the conventional group,and the difference was statistically significant(P<0.05);The number of fracture cycles(NCF)and the instrument flange length(FL)in the high-temperature steam sterilization group were higher than those in the conventional group,and the differences were statistically significant(P<0.05);The instruments in the high-temperature steam sterilization group mainly exhibited ductile cracking,accompanied by brittle intergranular cracking,clean and smooth grain boundaries,presenting flaky brittle precipitate phase,strong multi-faceted three-dimensional sense,and distinct edges.The conventional group showed ductile cracking,obvious micro-holes could be seen at the fracture,and the fracture surface was covered with a lot of dimple structures.Conclusion The disinfection effect of high-temperature steam sterilization is good,but the long-term use of high-temperature steam sterilization method for otolaryngological instruments causes more damage to the instruments than conventional wiping disinfection method.

关键词

高温蒸汽灭菌/耳鼻喉科/医疗器械/消毒效果/循环疲劳性能

Key words

High temperature steam sterilization/Otolaryngological/Instruments/Disinfection Effect/Cyclic fatigue performance