Quality Control Management Measures for Rigid Endoscopes Based on Maintenance Data Analysis
Objective To develop quality control management measures and ensure the safe use of rigid endoscopes,based on maintenance data analysis for rigid endoscope.Methods With the analysis of the maintenance data of 95 rigid endoscopes currently in use in the First Operating Room of the Affiliated Hospital of Xuzhou Medical University from January to December 2022,the root cause of the failure of the rigid endoscope was analyzed by investigating the responsible persons in various aspects such as the use,storage,and disinfection of the rigid endoscope,with the help of the fishbone diagram.Quality control management measures for rigid endoscopes,included systems,personnel,environment,and processes,were developed based on fishbone diagram analysis results.The maintenance data of rigid endoscopes before and after the implementation of quality control management measures were compared.Results After the implementation of quality control management measures,the maintenance times of rigid endoscopes decreased from 103 times to 69 times,with a decrease of 33.01%.The maintenance cost decreased from 551700 yuan to 371000 yuan,with a decrease of 32.75%.Conclusion The quality control management measures for hard endoscopes based on maintenance data analysis can effectively improve the management quality level of hospitals for rigid endoscopes and reduce the risk of use.
Rigid endoscopeQuality control managementFailureData analysis