首页|双源CT低剂量Turbo Flash扫描模式迭代重建在婴幼儿肺部感染中的应用价值

双源CT低剂量Turbo Flash扫描模式迭代重建在婴幼儿肺部感染中的应用价值

扫码查看
目的 探讨双源CT低剂量Turbo Flash扫描模式高级模拟迭代重建(ADMIRE)在婴幼儿非镇静状态下胸部CT检查中的图像质量、辐射剂量及临床应用价值.方法 前瞻性分析本院2022年3月至8月因临床怀疑肺部感染需要胸部CT检查的婴幼儿患者90例.随机分为100kV实验组、70kV实验组和常规扫描对照组,每组各30例.实验组均采用西门子第3代双源CT、Turbo Flash扫描模式,ADMIRE重建;常规扫描对照组采用西门子64层CT常规儿童胸部CT平扫协议.比较三组患儿的扫描长度、扫描时间、辐射剂量和图像质量.结果 100kV实验组的扫描时间、辐射剂量(SSDE)分别较常规扫描对照组下降88.48%、84.74%,差异有统计学意义(P均<0.01),较70kV实验组差异无统计学意义(P>0.05).3组图像均满足临床诊断要求,常规扫描对照组图像可见不同程度的运动伪影,图像质量评分均低于100kV实验组和70kV实验组,差异有统计学意义(P均<0.01).结论 婴幼儿胸部CT低剂量Turbo Flash扫描模式联合ADMIRE(3级)重建可在满足临床诊断的前提下明显缩短扫描时间减少运动伪影、大幅度降低患儿的辐射剂量、提高图像质量,保障婴幼儿患者非镇静状态胸部CT检查质量.
The Value of Dual-source CT Low-dose Turbo Flash Scan Pattern Iterative Reconstruction in Infantile Lung Infections
Objective To investigate the image quality,radiation dose,and clinical value of dual-source CT low-dose Turbo Flash scanning mode Advanced modeled iterative reconstruction(ADMIRE)in chest CT examinations of infants and children under non-sedation.Methods Prospective analysis of 90 infants and children who required chest CT examination for clinical suspicion of lung infection from March to August 2022 in our hospital.They were randomly divided into 100kV experimental group,70kV experimental group and conventional scanning control group,with 30 cases in each group.The experimental group all used Siemens 3rd generation dual-source CT,Turbo Flash scanning mode,and ADMIRE reconstruction;the routine scanning control group used Siemens 64-slice CT routine pediatric chest CT plain scanning protocol.The scan length,scan time,radiation dose and image quality of the children in the three groups were compared.Results The scanning time and radiation dose(SSDE)of the 100kV experimental group decreased by 88.48%and 84.74%,respectively,compared with that of the conventional scanning control group,with statistically significant differences(P<0.01),and the difference was not statistically significant compared with that of the 70kV experimental group(P>0.05).The images of the three groups met the requirements for clinical diagnosis,and the images of the conventional scanning control group showed different degrees of movement artifacts,and the image quality scores were lower than those of the 100kV experimental group and the 70kV experimental group.The quality scores of the three groups were lower than those of the 100kV experimental group and the 70kV experimental group,and the differences were statistically significant(P<0.01).Conclusion The combination of low-dose Turbo Flash scanning mode and ADMIRE(level 3)reconstruction of chest CT in infants and children can significantly shorten the scanning time and reduce the motion artifacts,substantially reduce the radiation dose to the children,improve the image quality,and guarantee the quality of non-sedated chest CT examination in infants and children,under the premise of meeting the clinical diagnosis.

InfantChestRadiation DoseTomographyX-ray ComputerIterative Reconstruction

孙升级、徐媛、李海萍、屈瑞、吴世杰、刘建莉

展开 >

兰州大学第二医院放射科(甘肃兰州 730030)

婴幼儿 胸部 辐射剂量 体层摄影术 X线计算机 迭代重建

甘肃省基础研究创新群体项目

21JR7RA432

2024

中国CT和MRI杂志
北京大学深圳临床医学院 北京大学第一医院

中国CT和MRI杂志

CSTPCD
影响因子:1.578
ISSN:1672-5131
年,卷(期):2024.22(7)
  • 5