中国介入心脏病学杂志2024,Vol.32Issue(1) :39-44.DOI:10.3969/j.issn.1004-8812.2024.01.008

不同对比剂浓度对准分子激光消蚀作用的影响:体外研究

The impact of different contrast agent concentrations on excimer laser ablation:an in vitro study

何攀 史杨 陈海威 杨俊杰 郭军 陈韵岱 汪奇
中国介入心脏病学杂志2024,Vol.32Issue(1) :39-44.DOI:10.3969/j.issn.1004-8812.2024.01.008

不同对比剂浓度对准分子激光消蚀作用的影响:体外研究

The impact of different contrast agent concentrations on excimer laser ablation:an in vitro study

何攀 1史杨 2陈海威 2杨俊杰 2郭军 2陈韵岱 2汪奇2
扫码查看

作者信息

  • 1. 中国人民解放军总医院第六医学中心心血管病医学部,北京 100048;中国人民解放军医学院,北京 100853
  • 2. 中国人民解放军总医院第六医学中心心血管病医学部,北京 100048
  • 折叠

摘要

目的 探讨对比剂浓度对准分子激光消蚀斑块作用的影响.方法 采用直径为0.9 mm的激光纤维导管,运用波长为308 nm的氯化氙准分子激光进行斑块模型消蚀实验.在准分子激光消蚀过程中,根据所注射对比剂浓度不同分为5组:生理盐水组、25%浓度组、50%浓度组、75%浓度组、100%浓度组.通过光学相干断层成像评估准分子激光消蚀后斑块管腔面积变化,以评估对比剂浓度对准分子激光消蚀作用的影响.同时,利用水听器测量准分子激光在不同对比剂浓度的液体中产生的冲击波压力,以探讨准分子激光的冲击波压力与其消蚀作用的相关性.结果 75%浓度组和100%浓度组的消蚀面积相近(P>0.05),均大于50%浓度组、25%浓度组和生理盐水组(均P<0.001).50%浓度组的消蚀面积显著大于25%浓度组和生理盐水组(均P<0.001),而25%浓度组则大于生理盐水组(P<0.001).对比剂浓度对准分子激光冲击波压力的影响也呈现相似的趋势.准分子激光产生的冲击波压力与其消蚀面积之间存在显著正相关(r=0.9987,P<0.001).结论 可以通过改变对比剂浓度来调节准分子激光对斑块组织消蚀作用的强度,为准分子激光联合对比剂注射技术在冠心病治疗中运用提供了指导.

Abstract

Objective To explore the impact of contrast agent concentration on the excimer laser's effect on plaque ablation.Methods Using a laser catheter with a diameter of 0.9 mm,we conducted plaque model ablation experiments employing a 308-nanometer xenon chloride excimer laser.During the excimer laser ablation process,five groups were formed based on the injected contrast agent concentrations:a saline group,25%concentration group,50%concentration group,75%concentration group,and 100%concentration group.Optical coherence tomography was utilized to assess the changes in plaque lumen area after excimer laser ablation,evaluating the impact of contrast agent concentration on the excimer laser's ablation efficacy.Simultaneously,a water manometer was used to measure the shockwave pressure generated by the excimer laser in liquids with different contrast agent concentrations,aiming to explore the correlation between the shockwave pressure of the excimer laser and its ablative effect.Results The ablation areas in the 75%concentration group and the 100%concentration group were similar(P>0.05),both exceeding those in the 50%concentration contrast agent group,25%concentration group,and saline group(all P<0.001).Specifically,the ablation area in the 50%concentration group was significantly larger than that in the 25%concentration group and saline group(both P<0.001),while the 25%concentration group was larger than the saline group(P<0.001).The influence of contrast agent concentration on the shockwave pressure of the excimer laser exhibited a similar trend.Additionally,there was a significant positive correlation between the shockwave pressure generated by the excimer laser and its ablation area(r=0.9987,P<0.001).Conclusions The intensity of excimer laser ablation on plaque tissue can be modulated by altering the contrast agent concentration.These findings offer guidance for the application of excimer laser in conjunction with contrast agent injection techniques in the treatment of coronary artery disease.

关键词

准分子激光/对比剂注射/消蚀作用/光学相干断层成像

Key words

Excimer laser/Contrast infusion/Ablation/Optical coherence tomography

引用本文复制引用

出版年

2024
中国介入心脏病学杂志
北京大学

中国介入心脏病学杂志

CSTPCDCSCD
影响因子:1.224
ISSN:1004-8812
参考文献量19