首页|国产BLT-D800血滤机与进口Braun Dialog+血滤机的临床应用对比

国产BLT-D800血滤机与进口Braun Dialog+血滤机的临床应用对比

Comparison of the application of domestic BLT-D800 and imported Braun Dialog+hemofiltration ma-chines

扫码查看
目的 研究Biolight D800(BLT-D800,广东宝莱特医用科技股份有限公司,中国)血滤机的性能.方法 以Braun Dialog+(贝朗爱敦股份公司,德国)血滤机为对照,采用随机对照法研究国产BLT-D800血滤机性能.选择2022年8月北京积水潭医院贵州医院血液透析中心接受规律血液透析治疗的MHD患者120例,采用随机数字法分为2组(60例/组),每周3次血液透析(hemodialysis,HD),每月1次后稀释血液透析滤过(hemodiafiltration,HDF),对比HDF治疗时血液透析并发症、治疗效果等指标和设备配液情况.结果 2组患者在HDF治疗过程中,低血压、高血压、肌肉痉挛、胸闷等不良反应发生率差异无统计学意义(x2=0.178,0.934,0.167,0.164,P=0.677,0.358,0.666,0.873);BLT-D800组HDF治疗后尿素清除指数(Kt/V)、尿素下降率(URR)、血β2微球蛋白(β2 microglobulin,β2-MG)等指标的改善程度与Braun Dialog+组比较差异无统计学意义(t=-0.426,0.165,0.046,P=0.381,0.692,0.833);超滤误差对照组更高,2组间差异有统计学意义(t=-14.956,P<0.001).结论 国产BLT-D800血滤机性能稳定,能安全有效开展HDF治疗.
Objective To investigate the performance of domestic Biolight D800 hemofiltration machine(BLT-D800 Guangdong Biolight Meditech CO.,China).Method This was a randomized controlled study,using the imported Braun Dialog+hemofiltration machine(B.Braun Avitum AG,Germany)as the control.A total of 120 patients undergoing maintenance hemodialysis(MHD)at the Blood Purification Center,Guizhou Hospital of Beijing Jishuitan Hospital were recruited for this study.They were divided into 2 groups(n=60/group)by a random number table method.Patients in each group were treated with hemodialysis(HD)3 times/week and post diluted hemodiafiltration(HDF)once/month.Dialysis complications,treatment effects,dialysate for the machine and other indicators were observed during HDF.Result The ratios of hypotension,hypertension,muscle cramp,and choking sensation in chest were similar between the two groups(x2=0.178,0.934,0.167 and 0.164 respectively;P=0.677,0.358,0.666 and 0.873 respectively).After HDF treatment,Kt/V,urea reduction rate(URR)and serum β2 microglobulin(β2-MG)were also similar between the two groups(t=-0.426,0.165 and 0.046 respectively;P=0.381,0.692 and 0.833 respectively).Ultrafiltration error was statisti-cally higher in the control group(t=-14.956,P<0.001).Conclusion The domestic BLT-D800 hemofiltration machine is stable for blood purification performance and is capable to carry out HDF safely and effectively

HemodiafiltrationHemofiltration machineDomestic

李琪、李晓颖、张庆华、张中琴

展开 >

550007 贵阳,北京积水潭医院贵州医院血液净化中心

血液透析滤过 血滤机 国产

2024

中国血液净化
中国医院协会

中国血液净化

CSTPCD
影响因子:1.54
ISSN:1671-4091
年,卷(期):2024.23(3)
  • 11