自旋锁定频率对心肌定量T1ρ成像的影响

Influence of Spin-Lock Frequency on Quantitative Assessment of Myocardial T1ρ Mapping

扫码查看
目的 探究在3.0T MR系统下,自旋锁定频率对心肌定量T1ρ值的影响.资料与方法 前瞻性纳入2023年7-9月在安徽医科大学第一附属医院行心脏磁共振检查的38名健康志愿者.使用3.0T MR系统进行T1ρ mapping及短轴电影成像扫描.获得自旋锁定频率为5 Hz、300 Hz、400 Hz、500 Hz的左心室短轴T1ρ图像,分别测得各节段T1ρ值和心肌纤维化指数,采用单因素重复测量方差分析观察不同自旋锁定频率下T1ρ值和心肌纤维化指数的差异.结果 5 Hz、300 Hz、400 Hz、500 Hz的T1ρ值分别为(33.9±2.8)ms、(43.4±2.1)ms、(45.4±2.6)ms、(46.5±2.4)ms,自旋锁定频率越高,T1ρ值越大,且不同自旋锁定频率的T1ρ值之间差异有统计学意义(300 Hz比 400 Hz:P<0.001;300 Hz比 500 Hz:P<0.001;400 Hz比 500 Hz:P=0.043).300 Hz、400 Hz、500 Hz下的心肌纤维化指数分别为(9.4±2.2)ms、(11.3±2.9)ms、(12.6±2.7)ms,差异有统计学意义(300 Hz比400 Hz:P<0.001;300 Hz比500 Hz:P<0.001;400 Hz比500 Hz:P=0.033).结论 本研究提供了特定心脏磁共振设置下健康成人的心肌T1ρ值,结果表明自旋锁定频率可以影响T1ρ值.
Purpose To explore the influence of different spin-locking frequencies on T1ρ values based on a 3.0T MR system.Materials and Methods Thirty-eight healthy adult volunteers underwent cardiac magnetic resonance imaging at the First Affiliated Hospital of Anhui Medical University from July to September 2023.T1ρ mapping and short-axis cine imaging with steady-state free precession sequences were performed with 3.0T MR system.T1ρ mapping sequence in three short-axis slices with three spin-lock frequencies at the amplitude of 5 Hz,300 Hz,400 Hz,and 500 Hz was scanned,respectively.T1ρ relaxation times and myocardial fibrosis index were quantified for each slice and each myocardial segment,the difference in T1ρ of different spin-locking frequencies and myocardial fibrosis index was analyzed using one-way repeated-measures analysis of variance method.Results T1ρ of 5 Hz,300 Hz,400 Hz,and 500 Hz were(33.9±2.8)ms,(43.4±2.1)ms,(45.4±2.6)ms and(46.5±2.4)ms,respectively;and T1ρ values showed a significant progressive increase from the low spin-lock frequency to the high spin-lock frequency of the heart(300 Hz vs.400 Hz:P<0.001;300 Hz vs.500 Hz:P<0.001;400 Hz vs.500 Hz:P=0.043).In addition,the measured myocardial fibrosis index at 300 Hz,400 Hz and 500 Hz were(9.4±2.2)ms,(11.3±2.9)ms and(12.6±2.7)ms,respectively.Statistical analysis underscored significant variations among these measurements(300 Hz vs.400 Hz:P<0.001;300 Hz vs.500 Hz:P<0.001;400 Hz vs.500 Hz:P=0.033).Conclusion In this prospective study,myocardial T1ρ values for the specific cardiac magnetic resonance setting are provided,and we found that spin-lock frequency can affect the T1ρ values.

MyocardiumMyocardial fibrosisMagnetic resonance imagingT1ρ mappingSpin-lock frequency

韩彩云、邓炜、赵韧、束宏敏、王震、杨金秀、余永强、李小虎

展开 >

安徽医科大学第一附属医院医学影像科,安徽 合肥 230022

安徽医科大学第一附属医院心血管内科,安徽 合肥 230022

心肌 心肌纤维化 磁共振成像 T1ρ mapping 自旋锁定成像

安徽省临床医学研究转化项目安徽省临床医学研究转化项目安徽省临床医学研究转化项目

202304295107020027202304295107020028202304295107020029

2024

中国医学影像学杂志
中国医学影像技术研究会

中国医学影像学杂志

CSTPCD北大核心
影响因子:1.37
ISSN:1005-5185
年,卷(期):2024.32(7)
  • 4