长板探测器数字X射线摄影术临床应用的可行性分析

Feasible analysis of DR long board detector in clinical application

扫码查看
目的:探讨长板探测器数字X射线摄影术(DR)应用于临床的可行性.方法:将长板探测器(简称探测器)直立竖放,金属铅线标记的刻度长尺立于探测器面板长轴中心前方20 cm处且与探测器中轴线平行.将3块空间分辨力测试方卡分别放置于探测器的上、中、下3个位置,互相之间相隔30 cm呈45°方位贴敷于探测器面板上.源图像距离(SID)分别为100、150和200 cm时曝光,检测探测器上、中、下3个选点位置不同SID时获得的入射剂量;通过对方卡的图像观察确定3个位置点的空间分辨力;通过对金属铅线标记的刻度长度的投影测量与实际铅线长度之比,获得不同点位的放大率.观察探测器长轴方向有效焦点面的空间分布情况及其形态变化.结果:当SID为100、150和200 cm时,图像放大率随SID增大而减小,3种SID放大率比较,差异有统计学意义(F=223.80,P<0.001);不同SID对应的辐射剂量比较,差异有统计学意义(F=7.57,P<0.05);空间分辨力恒定为1.8 LP/mm.探测器短轴方向存在足跟效应,探测器长轴方向有效焦点呈上下对称显示.结论:长板探测器DR成像设备实现了一次曝光摄取脊柱全长或下肢全长的影像,可满足临床需求,提高X射线检查效率.
Objective:To explore the feasibility of long board detector of digital radiography(DR)in clinical application.Methods:The long board detector(detector)was erected and placed upright.The scale long ruler with marked metal lead wire was placed at 20 cm in front of the center of long axis of the board of detector,which paralleled medial axis.Three test cards of spatial resolution were respectively placed at three positions(upper,middle and lower)of detector,and they were stuck on the board of detector as 30cm intervals between each other and 45° position.The exposures were conducted at 100,150,and 200 cm of source image distance(SID).The incident doses were tested,which obtained from different SID spots of upper,middle and lower positions of detector.The spatial resolutions of 3 positions were determined through observed the images of cards.The ratio of the marked scale length with metal lead wire to actual length of lead wire was measured through the projection of the scale length,so as to obtain the amplification rate of different spot positions.The spatial distribution of effective focal plane on the direction of long axis of detector,and the morphological change of that were observed.Results:When SID spots were respectively 100,150 and 200cm,the amplification rates of images decreased with increasing SID.The difference of amplification rates among three SID spots was significant(F=223.80,P<0.001).There was significant difference in the corresponding radiation doses among different SID spots(F=7.57,P<0.05).The spatial resolution was constantly 1.8 LP/mm.There was heel effect along with the direction of short axis of detector.The effective focal spot on the direction of long axis of detector appeared up-down symmetrical display.Conclusion:The long board detector of DR equipment has realized the capture for the images of the overall length of spine or the overall length of lower limbs in one exposure,which can meet the clinical requirement,and improve the detection efficiency of X-ray.

Long board detectorDigital radiography(DR)Distortion of image amplificationEffective focal spotSpatial distribution

王文海、贾绍田、殷广宁、宋俊峰、张立新、赵英威

展开 >

首都医科大学附属北京积水潭医院放射科 北京 100035

应急总医院放射科 北京 100028

北京贝特莱博瑞技术检测有限公司 北京 100013

中国医学科学院肿瘤医院放射科 北京 100021

首都医科大学附属北京友谊医院放射科 北京 101100

展开 >

长板探测器 数字X射线摄影术(DR) 影像放大失真 有效焦点 空间分布

2024

中国医学装备
中国医学装备协会

中国医学装备

CSTPCD
影响因子:0.882
ISSN:1672-8270
年,卷(期):2024.21(3)
  • 20