Philips iU22型及iE33型超声诊断设备故障维修案例分析
Six cases of ultrasonic maintenance of Philips iU22 and IE33
刘蕾 1沈亚军 1陈克龙 1闫瑾 1仲建生 1夏云成2
作者信息
- 1. 南通大学附属如皋医院如皋市人民医院医学装备科 如皋 226500
- 2. 南通大学附属医院如皋分院设备科 如皋 226500
- 折叠
摘要
总结Philips iU22和iE33型超声诊断设备的结构及常见故障,分析其6例故障原因,确定故障维修方案更换相应元器件或重新安装系统软件等方法维修技术路线,为医院同行的日常维修工作提供一些参考.
Abstract
The structure and common faults of Philips iU22 and iE33 ultrasound diagnostic equipment were introduced,the fault phenomena,causes and maintenance ideas in the daily work of this series of ultrasonic equipment were analyzed.The equipment was repaired after the failure occurred by analyzing the fault causes,replacing the corresponding components or reinstalling the system software etc.,so as to etc.,so as to provide a reference for the daily maintenance work of medical engineers.
关键词
超声诊断设备/模拟电压调制板/USB/HUB板/高压电路/通道板Key words
Ultrasound diagnostic equipment/Analog voltage modulation board/USB HUB board/High voltage circuit/Channel board引用本文复制引用
出版年