中国医学装备2024,Vol.21Issue(10) :194-197.DOI:10.3969/j.issn.1672-8270.2024.10.035

内窥镜液体膨腔泵校准方法研究

Research on calibration method of endoscopic liquid expansion pump

黄靖 段傲文 杨利 文晓波 朱海江 魏安海 龙恒宇 张和华
中国医学装备2024,Vol.21Issue(10) :194-197.DOI:10.3969/j.issn.1672-8270.2024.10.035

内窥镜液体膨腔泵校准方法研究

Research on calibration method of endoscopic liquid expansion pump

黄靖 1段傲文 1杨利 1文晓波 1朱海江 1魏安海 2龙恒宇 3张和华1
扫码查看

作者信息

  • 1. 陆军军医大学大坪医院医学工程科 重庆 400042
  • 2. 重庆大学微电子与通信工程学院 重庆 400044
  • 3. 石家庄医学高等专科学校口腔医学系 石家庄 050062
  • 折叠

摘要

针对内窥镜液体膨腔泵关键性能参数压力示值误差、流量示值误差和流量示值重复性提出对应的校准方法.选取医院在用的4台不同品牌的内窥镜液体膨腔泵进行校准,评估校准方法的可行性.在内窥镜液体膨腔泵的压力和流量量程范围内分别选取高、中、低3个校准点,每个流量点校准3次.校准结果显示,压力示值最大误差值4.3%,流量示值最大误差为9.3%,流量示值重复性最大值为0.8%,均满足《医用内窥镜 内窥镜功能供给装置液体膨腔泵》(YY/T0864-2011)和《医用注射泵和输液泵校准规范》(JJF1259-2018)最大允许误差的技术要求.内窥镜液体膨腔泵校准方法能完善该类设备的计量溯源体系,确保设备量值准确与可靠,提升产品质量,保障医疗安全.

Abstract

A corresponding calibration method was proposed for the key performance parameters of the endoscopic liquid expansion pump,such as the pressure indication error,the flow rate indication error and the flow rate repeatability.4 different brands of endoscopic liquid expansion pumps in clinical use in the hospital were selected for calibration,and the feasibility of the calibration method was evaluated.In the pressure and flow range of the endoscope liquid expansion pump,3 calibration points of high,medium and low were selected,and each flow point was calibrated 3 times.The calibration results showed that the maximum error of pressure indication was 4.3%,the maximum error of flow indication was 9.3%,and the maximum repeatability of flow indication value was 0.8%,all of which met the technical requirements for the maximum allowable error of"Medical Endoscopes.Endoscope Functional Supply Units.Irrigation Pump"(YY/T 0864-2011)and"Calibration Specification for Syringe Pumps and Infusion Pumps"(JJF 1259-2018).The calibration method for endoscopic liquid expansion pump can improve the metrological traceability system of this type of equipment,ensure the accuracy and reliability of equipment values,improve product quality,and ensure medical safety.

关键词

内窥镜液体膨腔泵/校准/示值误差/示值重复性

Key words

Endoscopic liquid expansion pump/Calibration/Indication error/Repeatability of indication