目的:基于MATLAB软件平台设计多层螺旋CT(MSCT)成像系统质量控制检测参数智能化测量程序,实现MSCT成像系统质量控制检测的智能化.方法:基于MATLAB软件的图形用户界面(GUI)功能,设计MSCT成像系统质量控制检测参数智能化测量程序(简称智能化测量程序).采用图像读取、二值化、基于霍夫变换的圆形检测等算法,对MSCT质量控制参数CT值(水)、噪声和均匀性进行自动测算,采用组内相关系数(ICC)分析人工测量方法与设计的智能测量程序检测结果之间的一致性.结果:智能化测量程序可自动评估MSCT成像系统质量控制检测参数CT值(水)、噪声和均匀性,且人工测量方法与智能化测量程序检测结果之间的一致性好(ICC值为0.881~0.985).结论:MSCT成像系统质量控制检测参数智能化测量程序可简化MSCT成像系统的质量检测参数测算流程,为MSCT成像设备质量控制提供可靠的检测工具,有效提高MSCT成像系统质量控制检测效率.
Objective:To design an intelligent measurement program of detection parameters of quality control of imaging system of multi-slice spiral computed tomography (MSCT) based on MATLAB software platform,so as to achieve intelligent detection for quality control of MSCT imaging system. Methods:We designed an intelligent measurement program for the detection parameters of quality control of MSCT imaging system (referred to as the intelligent measurement program) bases on the function of graphical user interfaces (GUI) of MATLAB software. A series of algorithms such as image reading,binarization,and circular detection based on the Hough transform were employed to conduct automatic measurement and calculation for CT values (water),noise and uniformity of the parameters of MSCT quality control. The Intraclass Correlation Coefficient (ICC) was adopted to analyze the consistency of the detection results between the manual measurement method and the designed intelligent detection program. Results:The designed intelligent measurement program in this study can automatically assess the detection parameters of quality control of the MSCT imaging system,which included CT values (water),noise and uniformity. There was favorable consistency in the detection results between the manual measurement method and the intelligent measurement program (range of ICC values was from 0.881 to 0.985). Conclusion:The intelligent measurement program of detection parameters of quality control of MSCT imaging system can simplify the process of calculating detection parameters of quality control of MSCT imaging system,and provide a reliable detection tool for quality control of MSCT imaging equipment,which can effectively improve the detection efficiency of quality control.

Multi-slice spiral computed tomography (MSCT)Quality controlIntelligent detectionPerformance parameter of CT

叶培筠、熊晖、陈坚、黄志峰、林亚妹、杨志杰、陈楚洁、杨米扬、洪成坤、张宇航、毛明惠、曾台鹏、付丽媛

展开 >

中国人民解放军联勤保障部队第九〇〇医院放射诊断科 福州 350025

福建中医药大学福总教学医院(第九〇〇医院)放射诊断科 福州 350025

中国融通医疗健康集团莆田九十五医院医学影像科 莆田 351100

多层螺旋CT(MSCT) 质量控制 智能检测 CT性能参数

2024

中国医学装备
中国医学装备协会

中国医学装备

CSTPCD
影响因子:0.882
ISSN:1672-8270
年,卷(期):2024.21(12)