Determination of related substances in tegoprazan by HPLC
Objective To determine the 3 related substances in tegoprazan by HPLC.Methods An Agilent Proshell 120 EC-C18(4.6 mm×100 mm,2.7 μm)column was used.The mobile phase A was 0.01 mol·L-1 ammonium acetate buffer,and the mobile phase B was acetonitrile,with linear gradient elution.The detection wavelength was 220 nm.The flow rate was 0.7 mL·min-1,and the column temperature was 40 ℃.The injection volume was 10 μL.Results The calibration curves of impurity A,B and C were 0.0972~1.944 μg·mL-1(r=0.9999),0.0509~1.018 μg·mL-1(r=0.9997),and 0.1089~2.178 μg·mL-1(r=0.9997).The average recovery was 96.27%,95.40%and 96.10%,with RSDs of 2.1%,2.3%and 2.2%.Conclusion The method is accurate and reliable,which can be used for the quality control of tegoprazan.